Specifications according to ISO 10360-2:2009, ISO 10360-5:2010 and ISO 10360-4:2000
| Probe Configuration | Temp 1 | erature Ran 8 ÷ 22 °C | ge | Temp 1 | erature R 6 ÷ 26 °C | ange | Temp 1 | erature R 5 ÷ 30 °C | ange |
| E0,MPE/E150,MPE | PFTU,MPE | MPETHP/t | E0,MPE/E150,MPE | PFTU,MPE | MPETHP/t | E0,MPE/E150,MPE | PFTU,MPE | MPETHP/t | |
| DEAGLOBAL Silver SF 09.xx.08 | |||||||||
| HP-S-X3C/X5 | 1.4 + L/333 | 1.4 | 2.5/45 | 1.4 + L/285 | 1.4 | 2.5/45 | 1.4 + L/256 | 1.4 | 2.5/45 |
| HH-A, HH-AS/HP-S-X1S, H | 1.4 + L/333 | 1.4 | 2.5/45 | 1.4 + L/285 | 1.4 | 2.5/45 | 1.4 + L/256 | 1.4 | 2.5/45 |
| HH-A, HH-AS/TP200 | 1.9 + L/333 | 1.9 | – | 1.9 + L/285 | 1.9 | – | 1.9 + L/256 | 1.9 | – |
| HH-A, HH-AS/HP-T, HP-TM | 2.1 + L/333 | 2.0 | – | 2.1 + L/285 | 2.0 | – | 2.1 + L/256 | 2.0 | – |
| DEAGLOBAL Silver SF 12.xx.10 | |||||||||
| HP-S-X3C/X5 | 2.1 + L/333 | 1.8 | 3.1/45 | 2.1 + L/277 | 1.8 | 3.1/45 | 2.1 + L/250 | 1.8 | 3.1/45 |
| HH-A, HH-AS/HP-S-X1S, H | 2.1 + L/333 | 1.8 | 3.1/45 | 2.1 + L/277 | 1.8 | 3.1/45 | 2.1 + L/250 | 1.8 | 3.1/45 |
| HH-A, HH-AS/TP200 | 2.5 + L/333 | 2.2 | – | 2.5 + L/277 | 2.2 | – | 2.5 + L/250 | 2.2 | – |
| HH-A, HH-AS/HP-T, HP-TM | 2.7 + L/333 | 2.6 | – | 2.7 + L/277 | 2.6 | – | 2.7 + L/250 | 2.6 | – |
| Probe Configuration | RO,MPL (1) | RO,MPL (2) | |||||||
| HP-S-X3C/X5 | 1.2 | 1.7 | |||||||
| HH-A, HH-AS/HP-S-X1S, H | 1.2 | 1.7 | |||||||
| HH-A, HH-AS/TP200 | 2.1 | 2.7 | |||||||
| HH-A, HH-AS/HP-T, HP-TM | 2.1 | 2.7 |
NOTE: E /E specifications are to be formally understood as E * /E * for the case where non-normal CTE material calibrated test lengths are used.
0, MPE 150, MPE 0,MPE 150,MPE
(1)R referred to DEA GLOBAL 09.xx.08 0,MPL (2)R referred to DEA GLOBAL 12.xx.10 0,MPL
Probe Configuration for performance test: Performance data are valid if the following specification are
| • HP-S-X1S: without extension, stylus length 50 mm, tip | Temperature Range | 18 ÷ 22 °C |
| diameter 5 mm | Max. air temperature variation | 1 °C/h - 2 °C/24h |
| • HP-S-X1H: with an extension of 100 mm, stylus length | Max. air temperature variation | 1 °C/m |
| 20 mm, tip diameter 5 mm | ||
| • TP200: Standard Force Module, stylus length 10 mm, tip | Temperature Range | 16 ÷ 26 °C |
| diameter 4 mm | Max. air temperature variation | 1 °C/h - 5 °C/24h |
| • HP-T: Standard Force, stylus length 10 mm, tip | Max. air temperature variation | 1 °C/m |
| diameter 4 mm | ||
| • HP-TM: Standard Force Module, stylus length 10 | Temperature Range | 15 ÷ 30 °C (*) |
| mm, tip diameter 4 mm | Max. air temperature variation | 1 °C/h - 5 °C/24h |
Max. air temperature variation 1 °C/m Dynamics:
2 2
Max. Scanning Speed : 500 mm/s Temperature Range 15 ÷ 30 °C
Max. air temperature variation 2 °C/h - 10 °C/24h Max. air temperature variation 1 °C/m